- 1.5X Magnification.
- Micro focus X-ray
tube.
- Automatic Exposure
Control (AEC).
- A molybdenum filter
(0.03mm Mo) and a Aluminum filter (0.5mm Al) are provided to absorb
unnecessary soft X-rays.
- Mo filter covers
low level kV range (22~35kV) and Al filter covers high kV range
(35~39kV) and Mo filter is useful for increasing image contrast
in large breast with large amounts of glandular tissue.
- Both motorized and
manual breast compression.
- The supporting C-Arm,
with motorized vertical travel of 820mm, is designed for complete
axial rotation (+/-180).
- ASP (Auto
Standard Exposure Positioning) makes operators easy to execute
4 axis exposures by software programming. This valuable software
controls the 4 standard positions (RMLO, RCC, LMLO, LCC)
of C-Arm automatically by one –touch operation
for MedioLateral Oblique view and CranioCaudal
view.
- Bucky device 18
x 24cm (or 24 x 30 cm option).
- Auto Film Changer.
|